3-27
CLEAR
PT
DATA
STORE
VS
DISP
VSC
PT
SETUP
HR
AUTO
ECG
TRACE
SEL
MORE
2. Press the CLEAR PT DATA softkey and press YES softkey.
3. Press the PAGE HOME key.
Oxycardiorespirogram (OxyCRG/CRG)
OxyCRG is an optional trend that allows display of compressed respiration waveform (1mm/sec)
correlated with trend data derived from heart rate and SpO2 over most recent two minutes of
monitoring. ECG, Respiration, and SpO2 must be actively monitored for proper monitoring of
OxyCRG. OxyCRG can be temporarily or continually displayed (Trace 2 only).
To temporarily display OxyCRG:
1. Monitor ECG, Respiration, and SpO2.
2. Press the SYSTEM SETUP key.
CLEAR
PT
DATA
STORE
VS
DISP
VSC
PT
SETUP
HR
AUTO
ECG
TRACE
SEL
MORE
3. Press the MORE softkey to display the following menu.
ADULT
PED
NEO
VOL
DISP
OCRG
REC
SETUP
TIMER CLOCK MORE
4. Press the DISP OCRG to display OxyCRG.