3-21
1. Inspect circular window on airway adapter for cracks or blemishes. Replace if necessary.
2. Attach airway adapter and sensor in line proximally to patient between endotracheal,
nasotracheal or tracheal tube and ventilator circuit, T-tube setup or bag/valve setup.
3. Carefully press sensor down onto airway adapter being careful not to damage glass window.
4. Connect ETCO2 cable to port on monitor. Function automatically activates.
5. Allow device to warm up for 45-60 seconds. The message WARM-UP is present on screen.
6. Support under sensor housing may be necessary to prevent excess traction on airway tube. Do
not allow sensor housing to set on patient’s skin.
Sidestream:
The sidestream CO2 sensor attaches to a nasal cannula or in-line airway sampling system (for
intubated patients). It can even attach to a divided oxygen delivery CO2 sampling nasal cannula if
you need to deliver O2 while monitoring sidestream CO2.
1. Connect ETCO2 cable to monitor connector.
2. Insert fresh watertrap into sidestream motor port with a gentle twisting motion.
3. Place respiratory device (nasal cannula, facemask, airway tubing) to patient and attach
sampling port line from respiratory device to watertrap.
4. Allow device to warm up for 45-60 seconds. The message WARM-UP is present on screen.
Vedere la pagina 61