3-7
MODE for Nellcor MP405 Averaging Time Application
*NORM (NOT USED) (NOT USED) (NOT USED)
FAST (NELLCOR Mode 2) Typically 7-8 seconds for
average pulse.
Varies + – with pulse quality
and other functions.
All monitoring situations.
*SLOW (NOT USED) (NOT USED) (NOT USED)
*Nellcor OxiSmartXL Technology has eliminated the need for operation in NORM and
SLOW modes. Refer to the following NOTE.
NOTE: Since Nellcor OxiSmartXL technology has eliminated the need
for operation in the SLOW mode, and depending on the version of
main software, the following may occur if the SLOW mode is
accidentally selected:
Pulse waveform will be replaced by dotted lines
Dashed lines will replace pulse rate
Dashed lines will replace saturation values
A“02 WRT ER” message will be posted on the display
Upon returning to NORM or FAST mode, all SPO2 values and
waveform will return, and the “02 WRT ER” message will disappear
from the display.
WARNING: Incorrect sensor application or use may cause tissue
damage or improper operation of the monitor. Follow directions for
use provided with sensor.
NOTE: Regarding use of NELLCOR sensors:
Before use, carefully read NELLCOR sensor directions for use.
Use only NELLCOR oxygen transducers. Use of oxygen transducers
produced by other manufacturers may result in improper
performance.
Do not use damaged NELLCOR sensors or patient cables. Do not
use a NELLCOR sensor with exposed optical components. Do not
immerse sensor or patient cable in water, solvents, or cleaning
solutions (sensors, cable, and connectors are not waterproof). Do
not sterilize by irradiation, steam, or ethylene oxide. See cleaning
instructions in Directions for Use for reusable NELLCOR sensors
and patient cables.