12-4
Pulse Tone: Pitch varies with SpO
2
value
ECG Sync: NELLCOR C-LOCK
SpO
2
(Nellcor MP405)
Display: % SpO
2
, Plethysmographic Waveform or Pulse Amplitude Bar
Range: 0 to 100%
Accuracy: ADULT: 0 to 69% unspecified; 70 to 100% ±2 digits; NEO: 0 to 69%
unspecified; 70 to 100% ±3 digits
Pulse Rate Range: 20 to 250 BPM
Pulse Rate Accuracy: ±3 BPM
High Limit Range: 70 to 100%
Low Limit Range: 50 to 99%
Sensors: Neonatal through Adult (NELLCOR)
Pulse Tone: Pitch varies with SpO
2
value
ECG Sync: NELLCOR C-LOCK
SpO
2
(Masimo SET)
Display: % SpO
2
, Plethysmographic Waveform or Pulse Amplitude Bar
Range: 1 to 100%
Accuracy:
Calibrated to functional saturation
During NO Motion Conditions:
Saturation (% SpO
2
):
ADULT, PED: 70 to 100% ±2 digits
NEO: 70 to 100% ±3 digits
Pulse Rate: 25 to 240 bpm ±3 digits (Adult, Ped, and Neo)
During Motion Conditions:
Saturation (% SpO
2
):
ADULT, PED: 70 to 100% ±3 digits
NEO: 70 to 100% ±3 digits
Pulse Rate: 25 to 240 bpm ±5 digits (Adult, Ped, and Neo)
Low Perfusion Performance:
>0.02% Pulse Amplitude Saturation (%SpO
2
)±2 digits and %Transmission > 5%
Pulse Rate (bpm)±3 digits
High Limit Range: 70 to 100%
Low Limit Range: 50 to 99%
Sensors: Neonatal through Adult (Masimo)
Pulse Tone: Pitch varies with SpO
2
value
INVASIVE PRESSURE
Channels: 1, 2, or 3
Range: -40 to 300 mmHg
Accuracy: ±1% or ±1 mmHg
Parameter Display: SYS, DIAS, and MEAN selectable
Pressure Labels: ART, PA, CVP, ICP, RA, LAP, FEM, UA, UV, PRES
Alarm Limit Range:
Systolic: ADULT 0-240 mmHg
PED 0-210 mmHg
NEO 0-210 mmHg
Diastolic: ADULT 0-180 mmHg
Commenti su questo manuale