1-13
1. External DC Input (12 - 28V) Connector
Allows an external DC power source to be connected for extended use during transport where AC
power is not available. Batteries are not charged with external DC power.
2. Equipotential Connector (Grounding Terminal)
Provides grounding for the monitor when the monitor is used with other medical equipment. It
must be used as a protective ground terminal when the monitor is operated with external DC
power.
3. Batteries
These two slots contain modular batteries which provide power in transport situations or when
AC power is not available.
4. Defibrillator/Pacer Interface
Provides connection and communication between the Prism monitor and the ESCORT
Defibrillator/Pacer.
5. Fuse Holder/Voltage Selector and Fuse Replacement
The fuse holder contains the fuses and the voltage selector for the monitor. Open the door of this
holder to replace fuses or to configure the voltage selection block — 115 VAC or 230 VAC. The
present voltage setting can be read through the small rectangular window on the door of the
holder. It is recommended that qualified technical service personnel replace fuses or change the
voltage selection when necessary.
WARNING: For protection against fire, replace the fuses only with
those of the same type and rating.
The Prism requires two identical slow blow 0.5A fuses for 115 VAC operation, or two slow blow
0.25A fuses for 220 VAC operation. Ensure that the AC power cord has been disconnected before
replacing fuses. Carefully open the door of the fuse holder with a short 1/8" flat screwdriver.
Replace the blown or defective fuse with one of the same type and rating. Noting orientation,
gently slide the fuse holder back until it locks snugly into its original place. After the fuse
replacement, connect the AC power cord to the Prism. The unit is now ready for operation.
6. AC Power Cord Input Connector and Optional Power Cord Mounting Bracket
Allows connection of the AC power cord to the monitor. In addition, the AC power cord must be
plugged into an AC source before the monitor can operate on AC power and before the batteries
can be charged.
For permanent connection, the MDE power cord bracket should be installed. See the following
figure for instructions on installing the power cord mounting bracket.